
6 Defibrillator/Transcutaneous Pacer Analyzers
Specications
Defibrillator Analyzer
Energy output measurement Compatible defibrillator waveshapes: Lown, Edmark, trapezoidal, dc bi-phasic, and ac pulsed bi-phasic
Autoranged measurement 0.1 J to 600 J
Accuracy 0.1 J to 360 J: ± 1 % of reading +0.1 J
360 J to 600 J: ± 1 % of reading +0.1 J, typical
Note: For pulsed bi-phasic defibrillator, specified accuracy is ± (1.5 % of reading + 0.3 J) on both ranges
Load resistance
Resistence: 50 Ω
Accuracy 1 %, non-inductive (< 2 μH)
Charge time measurement Range: 0.1 s to 100 s
Accuracy: ± 0.05 s, typical
Synchronization test
(cardioversion)
Delay time measurement
• Timing window: ECG R-wave peak to the defib pulse peak
• Range: -120 ms to 380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms
following the R-wave peak
Automated defibrillator test ECG waves
• Normal sinus: 10 BPM to 300 BPM in 1 BPM steps
• Ventricular fibrilation: Coarse and fine
• Monomorphic ventricular tachycardia: 120 BPM to 300 BPM in 1 BPM steps
• Polymorphic ventricular tachycardia: 5 types
• Asystole
ECG waves
ECG general Lead configuration: 12-lead simulation; RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance
1000 Ω
Rate accuracy ± 1 % nominal
ECG amplitudes Reference lead: Lead II (default) or Lead I
Settings: 0.05 mV to 0.45 mV by 0.05 mV and 0.5 mV to 5 mV by 0.05 mV
Accuracy: ± 2 % of setting (Lead II), ± 5 % for all other leads and defib paddles
ECG normal sinus Rates: 10 BPM to 360 BPM in 1 BPM steps
ECG on defibrillatorinput load Same as the Lead II amplitude but limited to ± 4 mV
ECG performance waves Square wave: 2 Hz and 0.125 Hz
Triangular wave: 2 Hz and 2.5 Hz
Sine waves: 0.05 Hz, 0.5 Hz, 5 Hz, 10 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, 150 Hz, and 200 Hz
Pulse: 30 BPM and 60 BPM, 60 ms pulse width
R-wave detection Waveform: Haver-triangle
Rate: 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Widths: 8 ms, 10 ms, 12 ms, and 20 ms to 200 ms in 10 ms steps
Accuracy: ± 1 % setting 0.2 mV
Noise immunity Wave sine
Line frequency: 50 Hz or 60 Hz (± 0.5 Hz)
Amplitude: 0 mV to 10 mV (by 0.5 mV ± 5 %)
Arrhythmia selections Pacer interactive (Impulse 7000DP only)
Supraventricular
Premature
Ventricular
Conduction
Transveneous paced with selectable pacer spike amplitudes and widths
Impulse 6000D/7000DP
Debrillator/Transcutaneous Pacemaker Analyzer
riešenia na presné meranie
Komentarze do niniejszej Instrukcji