32 Multi-Function Simulators
Key features
All-in-one complete monitor
testing 80 % smaller and
17 lbs/7.7 kilos lighter than
predecessor technology
8-in-1 multifunction simulator
tests ECG (including fetal ECG
and arrhythmias), respiration,
temperature, IBP, cardiac
output/cardiac catheterization,
NIBP, SpO2, and Rainbow
multi-wavelength waveforms
Stay-connected ECG posts for
easy/secure ECG snap and lead
connections
Custom SpO2 r-curve for
accurate testing of the
latest and future oximetry
technologies
Static pressure linearity testing
Repeatable NIBP simulation
(+/- 2 mmHg) for dynamic
pressure repeatability testing
Physiologically synchronized
pulses across all parameters
Barcode scanning and direct
data capture and printing
functionality
Onboard, customizable patient
pre-sets and autosequences for
fast/easy testing
Multi-language user interface
offers choice of language
selection
Integrated, easily-replaceable
long-life battery
Optional PC-interface software
offers customizable procedures/
checklists to replace bulky
service manuals and automated
data capture/storage*
Wireless communication
for remote PC control of test
device, as well as data transfer
and automated regulatory
reporting*
*You must have Ansur Test Executive
version 2.9.6 or greater on your PC to
communicate with the product
The 8-in-1 ProSim 8
Vital Signs Simulator
offers fast and compre-
hensive preventative
maintenance (PM)
testing for your entire
patient monitor fleet.
Designed to get you
in and out of most PM
locations in minutes,
this multifunction
simulator tests ECG
(including fetal ECG
and arrhythmias),
respiration, tempera-
ture, IBP, cardiac output/cardiac catheterization, NIBP, SpO
2
, and is capable
of testing Rainbow multi-wavelength waveforms. Featuring specialized
stay-connected ECG posts for secure lead connections, physiologically-
synchronized pulses across all parameters, and customizable patient pre-
sets and autosequences, the ProSim 8 patient simulator provides unbeatably
fast and easy complete monitor testing. Barcode-scanner compatibility and
wireless PC interface, direct printing, data transfer and reporting, along
with advanced and integrated technologies allow top confidence in patient
monitor fleet performance and supports passing regulatory audits with ease.
Specications
ProSim 8
Vital Signs Simulator
Normal-sinus-rhythm waveform
ECG reference The ECG amplitudes specified are for Lead II (calibration), from the base-
line to the peak of the R wave. All other leads are proportional
Normal sinus
rhythm
12-lead configuration with independent outputs referenced to right leg (RL).
Output to 10 universal ECG jacks, color-coded to AHA and IEC standards
High-level output 0.5 V/mV ± 5 % of the ECG amplitude setting available on a BNC connector
Amplitude 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5 mV (0.25 mV steps)
Amplitude accuracy ± (2 % of setting + 0.05 mV)
ECG rate 10 BPM to 360 BPM in 1 BPM steps
Rate accuracy ± 1 % of setting
ECG waveform
selection
Adult (80 ms) or pediatric (40 ms) QRS duration
ST-segment
elevation
Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps).
Additional steps: + 0.05 mV and - 0.05 mV
Power-on default 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Pacemaker waveform
Pacer pulse Amplitude 0 (off), ± 2, ± 4, ± 6, ± 8, ± 10, ± 12, ± 14, ± 16, ± 18, ±
20, ± 50, ± 100, ± 200, ± 500, and ± 700 mV for lead II
(reference lead)
Accuracy Reference lead II: ± (5 % setting + 0.2 mV)
All other leads: ± (10 % setting + 0.4 mV)
Pacer pulse width 0.1 ms, 0.2 ms, 0.5 ms, 1 ms, and 2 ms ± 5 %
Paced arrhythmias Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-ventricular sequential
Noncapture (one time)
Nonfunction
Power-on default Amplitude 5 mV, width 1 ms, atrial waveform