Electrical Safety Analyzers
Key features
• Superior compliance with
multiple standards: IEC60601-
1(partial), IEC62353, VDE 751,
ANSI/AAMI ES1:1993, NFPA-99,
AN/NZS 3551, IEC61010
• Three test loads
• Expanded leakage ranges
through 10,000 µA
• Dual-lead resistance, leakage,
and voltage tests
• AC only, dc only and true-rms
leakage readings
• 100 % and 110 % mains volt-
age for mains on applied part
(lead isolation) test
• DSP filter technology for
improved accuracy in leakage
measurements
• 20 A equipment current
• More applied parts selections
• ECG and performance
waveforms
• Intuitive user interface
• Easy-to-use applied parts (ECG)
connections
• Insulation posts on applied
parts connections
• Five different insulation tests
• Varying insulation test voltage
500 V dc and 250 V dc
• 2- or (optional) 4-wire ground
wire resistance
• Large display with adjustable
contrast
• Ergonomic design
• Optional Ansur plug-in
software
• USB connection
• CE, C-TICK and CSA for USA
and Canada
• RoHS compliance
• Optional Ansur test automation
software to standardize testing
procedures, capture waveforms,
and print and document test
result
The ESA620 Electrical
Safety Analyzer repre-
sents the next generation
in manual, portable elec-
trical safety testers. With
selections of three test
loads, protective earth
test currents, and two
insulation test voltages
this versatile product
can be used worldwide
to enhance productivity
and test to standards of
choice.
New DSP technology offers better accuracy of leakage measurements
throughout the ranges specified in the standards.
Equipped with ten safety-enhanced ECG posts, the ESA620 offers
simulation of ECG and performance waveforms so both electrical safety
and basic tests on patient monitors can be performed with a single con-
nection. When used with optional Ansur computer-based software plug-in,
the ESA620 becomes automated. This allows for standardization of test
procedures, capturing and storage of results, comparison to standard limits,
and printing of reports thus enabling the sophisticated performance of the
high-end electrical safety analyzers.
Voltage
Range (mains voltage) 120 V model: 90 V ac to 132 V ac rms
230 V model: 180 V ac to 264 V ac rms
Accuracy ± (2 % of reading + 1 V)
Range (accessible voltage) 0 V ac to 300 V ac rms
Accuracy ± (2 % of reading + 2 LSD)
Voltage tests Mains, Accessible, and Point to Point
Earth resistance
Modes Two terminal or four terminal
Test current > 200 mA ac or 10 A ac to 25 A ac
Ranges
0 Ω to 2 Ω
Accuracy
± (2 % of reading 0.015 Ω)
Equipment current
Mode AC rms
Range 0 A to 20 A
Accuracy ± 5 % of reading ± (2 counts or 0.2 A, whichever is greater)
Leakage current
Patient load selection
(input impedance)
AAMI ES1-1993 Fig 1
IEC 60601: Fig 15
IEC 61010: Fig A-1
Crest factor ≤ 3
Ranges 0 μA to 199.9 μA
200 μA to 1999 μA
2.0 μA to 10.0 mA
Frequency response DC to 1 kHz
1 kHz to 100 kHz
100 kHz to 1 MHz
Accuracy ± (1 % of reading + 1 μA or 1 LSD, whichever is greater)
± (2 % of reading + 1 μA or 1 LSD, whichever is greater)
± (5 % of reading + 1 μA or 1 LSD, whichever is greater)
ESA620
Electrical Safety Analyzer
Specications