Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 35
Temperature Test
Tools Needed IBP Temp Simulator, or equivalent
Test Setup None
Do this... Observe This... Pass/Fail
1. Connect the temperature simulator to the
Propaq MD T1 channel.
2. On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
oo
3. Repeat steps 1 and 2 of this procedure for the
T2 channel.