Strona 1 - MD Service Manual

Propaq® MD Service Manual9650-0801-01 Rev. ASeptember 2010

Strona 2

CHAPTER 1MAINTENANCE TESTS6 9650-0801-01 Rev. AThis chapter describes the following maintenance tests:Calibration and Calibration Checks• NIBP Calibra

Strona 3 - Table of Contents

Calibration and Calibration Checks9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 7Calibration and Calibration ChecksNIBP Calibration CheckEquipment

Strona 4 - Appendix B

CHAPTER 1MAINTENANCE TESTS8 9650-0801-01 Rev. AAcceptable ResultsThe unit passes the test if Pd-Primary is less than or equal to 3.0 mmHg and Pd-Safet

Strona 5 - Overview

Calibration and Calibration Checks9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 9CO2 Calibration Procedure (required for annual calibration)Equipm

Strona 6 - Safety Considerations

CHAPTER 1MAINTENANCE TESTS10 9650-0801-01 Rev. ACO2 Calibration Check ProcedureEquipment 5% CO2 calibration gas cylinder with CO2 sample line (kit ava

Strona 7 - Technical Service

NIBP Valve, Pump, and Tubing Integrity Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 11NIBP Valve, Pump, and Tubing Integrity TestsNIBP Leak

Strona 8 - 4 9650-0801-01 Rev. A

CHAPTER 1MAINTENANCE TESTS12 9650-0801-01 Rev. ACorrective ActionIf the unit fails the test, the most likely cause is damaged or disconnected internal

Strona 9 - Maintenance Tests

NIBP Valve, Pump, and Tubing Integrity Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 13 NIBP Inflation TestEquipment Single lumen BP hose, a

Strona 10 - Functional Verification Tests

CHAPTER 1MAINTENANCE TESTS14 9650-0801-01 Rev. AOperational TestsThe Propaq MD comes equipped with routine operational tests that can be performed per

Strona 11 - NIBP Calibration Check

Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 15Accessing the Operational TestsThe operational tests on the unit can be accessed

Strona 12

ZOLL is a registered trademark, and SurePower is a trademark of ZOLL Medical Corporation. Propaq is a registered trademark of Welch Allyn, Inc. Nellco

Strona 13 - Procedure

CHAPTER 1MAINTENANCE TESTS16 9650-0801-01 Rev. AKeypad TestThis test checks all the front panel buttons on the unit to make sure they are working corr

Strona 14 - Calibration Check Procedure

Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 17LCD TestThis test checks the colors of the Liquid Crystal Display on the unit.Fol

Strona 15 - NIBP Leak Test

CHAPTER 1MAINTENANCE TESTS18 9650-0801-01 Rev. AAudio TestThis test checks the audio tones on the unit to make sure they are working correctly.Follow

Strona 16 - NIBP Deflation Test

Operational Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 19Battery TestThis test checks the battery to make sure it is working correctly.Fol

Strona 17 - NIBP Inflation Test

CHAPTER 1MAINTENANCE TESTS20 9650-0801-01 Rev. ANIBP TestThis allows the operator to perform various checks of the NIBP module to make sure it is work

Strona 18

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 21Functional Verification TestsBefore You Begin the Functional Verifica

Strona 19

CHAPTER 1MAINTENANCE TESTS22 9650-0801-01 Rev. A Leads TestTools Needed 3-lead, 5-lead, and 12-lead cablesTest each cable separately.Test Setup Connec

Strona 20 - LED Test

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 23Heart Rate Display TestTools Needed ECG simulatorECG Cable (3-lead, 5

Strona 21

CHAPTER 1MAINTENANCE TESTS24 9650-0801-01 Rev. AHeart Rate Alarm TestTools Needed ECG simulatorDo this... Observe this... Pass/Fail1.Press the power b

Strona 22 - Printer Test

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 2510. On the Propaq MD, press the Alarm Suspend key ( ).Alarm is enable

Strona 23

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual iTable of ContentsPrefaceOverview...

Strona 24 - NIBP Test

CHAPTER 1MAINTENANCE TESTS26 9650-0801-01 Rev. ALeakage Current TestTools Needed See the manufacturer’s instructions or supplied specifications for th

Strona 25

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 27Shock TestTools Needed Fluke Impulse 7000 DP Analyzer, or equivalentT

Strona 26 - Leads Test

CHAPTER 1MAINTENANCE TESTS28 9650-0801-01 Rev. ASynchronized Cardioversion TestTools Needed Fluke Impulse 7000DP or equivalent defibrillator analyzer.

Strona 27 - Heart Rate Display Test

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 29Pacer TestTools Needed Fluke Impulse 7000 DP Analyzer, or equivalentT

Strona 28 - Heart Rate Alarm Test

CHAPTER 1MAINTENANCE TESTS30 9650-0801-01 Rev. APrinter TestTools Needed NoneTest Setup NoneDo this... Observe this... Pass/Fail1.On the Propaq MD, in

Strona 29

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 31SpO2 TestTools Needed SpO2 Simulator: Fluke Index 2 or equivalent, Ne

Strona 30 - Leakage Current Test

CHAPTER 1MAINTENANCE TESTS32 9650-0801-01 Rev. AEtCO2 TestTools Needed CO2 sampling line airway adapterTest Setup NoneDo this... Observe this... Pass/

Strona 31 - Shock Test

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 33IBP TestTools Needed IBP patient simulator (with accuracy better than

Strona 32

CHAPTER 1MAINTENANCE TESTS34 9650-0801-01 Rev. ANIBP TestTools Needed NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinderTest

Strona 33 - Pacer Test

Functional Verification Tests9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 35Temperature TestTools Needed IBP Temp Simulator, or equivalentTest Se

Strona 34

TABLE OF CONTENTSii 9650-0801-01 Rev. AChapter 4 Life Cycle CountersOverview...

Strona 35

CHAPTER 1MAINTENANCE TESTS36 9650-0801-01 Rev. A

Strona 36

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 37Chapter 2TroubleshootingOverviewThis chapter contains a list of error messages that users may see

Strona 37 - IBP Test

CHAPTER 2TROUBLESHOOTING38 9650-0801-01 Rev. AQUICK ACCESS SW 1 SHORTQuick access key 1 is stuck. Replace the front panel.QUICK ACCESS SW 2 SHORTQuick

Strona 38

Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 39PS DEFIB CHARGE SHORTPaddle set keys are stuck. Replace the paddle set or

Strona 39 - Temperature Test

CHAPTER 2TROUBLESHOOTING40 9650-0801-01 Rev. AIBP DEVICE FAILURE IBP software detected a problem with the IBP data that is most likely a hardware issu

Strona 40

Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 41NIBP COMMUNICATION FAILUREThere is a communication problem with the NIBP m

Strona 41 - Troubleshooting

CHAPTER 2TROUBLESHOOTING42 9650-0801-01 Rev. ANIBP AIR LEAK Major air leak or could not reach target pressure.Attempt to identify problems with the ho

Strona 42 - CHAPTER 2TROUBLESHOOTING

Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 43NIBP NEONATAL DETECTEDDevice thinks a neonate cuff is connected. If no neo

Strona 43 - Propaq MD Error Messages

CHAPTER 2TROUBLESHOOTING44 9650-0801-01 Rev. ACO2 GAS INPUT OCCLUSIONCO2 inlet blocked.Replace FilterLine®. If problem persists, check for CO2 inlet o

Strona 44

Propaq MD Error Messages9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 45PS DEFIB ENERGY DOWN SHORTPaddle set energy down select button is stuck.Re

Strona 45

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 1PrefaceOverviewZOLL Medical Corporation’s Propaq MD Service Manual is intended for the trained biom

Strona 46

CHAPTER 2TROUBLESHOOTING46 9650-0801-01 Rev. A

Strona 47

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 47Chapter 3Disassembly ProceduresOverviewThis chapter provides instructions on how to disassemble an

Strona 48

CHAPTER 3DISASSEMBLY PROCEDURES48 9650-0801-01 Rev. A• Kapton tape• 3M copper adhesive tape, or equivalent• 1/2” nut driver• Large diagonal cutters• S

Strona 49

Removing the Handle9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 49Removing the HandleTools Required• #1 Phillips screwdriver with torque capabili

Strona 50

50 9650-0801-01 Rev. ARemoving the Printer ModuleTools Required• #1 Phillips screwdriver with torque capability• Orange (wooden) stickSetup• This proc

Strona 51 - Disassembly Procedures

Removing the Printer Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 516 Disconnect the two (2) cable connectors from the printer module.conne

Strona 52 - Safety Precautions

52 9650-0801-01 Rev. ADisassembling the Main UnitTool Required• # 1 Phillips screwdriver with torque capability• orange (wooden) stickSetup• This proc

Strona 53 - Removing the Handle

Disassembling the Main Unit9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 536 Disconnect the battery communication cable (yellow, orange, brown, re

Strona 54 - Removing the Printer Module

54 9650-0801-01 Rev. A11 Separate the two parts of the unit.

Strona 55

Removing the Front Enclosure Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 55Removing the Front Enclosure ModuleTools Required• #2 Phillips

Strona 56 - Disassembling the Main Unit

PREFACE2 9650-0801-01 Rev. AChapter 6—Functional Description provides technical descriptions for the Propaq MD major subassembly modules.Appendix A—Pr

Strona 57

56 9650-0801-01 Rev. A6 Remove the ZIF (Zero Insertion Force) flex cable using an orange stick.

Strona 58 - 54 9650-0801-01 Rev. A

Removing the Display Module9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 57Removing the Display ModuleTools Required• #1 Phillips screwdriver with

Strona 59 - Tools Required

58 9650-0801-01 Rev. A5 Flip the unit back over so the display screen is facing up.6 Lift the casting up while pushing down on the rubber boot that ru

Strona 60 - 56 9650-0801-01 Rev. A

Removing the EtCO2 and NIBP Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 59Removing the EtCO2 and NIBP ModulesTool Required• # 1 Phillips

Strona 61 - Removing the Display Module

60 9650-0801-01 Rev. A3 Disconnect multi-wire IBP cable with ferrite in the area illustrated.4 Cut the tie wrap to the IBP cable in the area shown wit

Strona 62 - 58 9650-0801-01 Rev. A

Removing the EtCO2 and NIBP Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 619 Remove five (5) screws and washers from the NIBP unit to rele

Strona 63 - and NIBP Modules

62 9650-0801-01 Rev. A14 Tilt side panel to access the EtCO2 connector and use an orange stick release the tabs and remove the EtCO2 connector.15 Remo

Strona 64 - 60 9650-0801-01 Rev. A

Removing the Monitor Board and SpO2 Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 63Removing the Monitor Board and SpO2 ModulesTool Require

Strona 65

64 9650-0801-01 Rev. A3 Lift the monitor board to expose the multi-wire LCD cable on the back side of monitor board and disconnect it.4 Remove three (

Strona 66 - 62 9650-0801-01 Rev. A

Removing the Defibrillator Board and Capacitor Bank Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 65Removing the Defibrillator Board and Ca

Strona 67 - Modules

Conventions9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 3Conventions WARNING! Warning statements describe conditions or actions that can result i

Strona 68 - 64 9650-0801-01 Rev. A

66 9650-0801-01 Rev. A5 Remove screw in the middle of the plastic isolator (located on top of the processor board) in the area illustrated.Note: Durin

Strona 69

Removing the Defibrillator Board and Capacitor Bank Modules9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 6711 Locate the black ferrite. Using the

Strona 71

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 69Chapter 4Life Cycle CountersOverviewThis chapter contains instructions for setting life cycle coun

Strona 72 - 68 9650-0801-01 Rev. A

CHAPTER 4LIFE CYCLE COUNTERS70 9650-0801-01 Rev. ACO2On the Propaq MD, do the following:1. Press the Setup quick access key ( ).2. Use the navigation

Strona 73 - Life Cycle Counters

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 71Chapter 5 Replacement PartsOverviewThis section contains a listing of the replacement parts availa

Strona 74 - Defibrillator/Pacer

CHAPTER 5REPLACEMENT PARTS72 9650-0801-01 Rev. AReplacement PartsPART NUMBER QUANTITY DESCRIPTION705519 1 Handle Top705517 1 Handle Bottom705510 4 Scr

Strona 75 - Replacement Parts

Replacement Parts9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 737772-0001-01 1 Printer Assembly7772-0002-01 1 Front Enclosure Assembly7772-0003-0

Strona 76 - Replacement Parts

CHAPTER 5REPLACEMENT PARTS74 9650-0801-01 Rev. A

Strona 77

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 75Chapter 6Functional DescriptionOverviewThis chapter contains functional descriptions of the compon

Strona 78 - CHAPTER 5REPLACEMENT PARTS

PREFACE4 9650-0801-01 Rev. ATechnical Service for International CustomersInternational customers should return the unit in its original container to t

Strona 79 - Functional Description

CHAPTER 6FUNCTIONAL DESCRIPTION76 9650-0801-01 Rev. AMain Processor (MP)The MP section controls the user interface, connectivity needs, as well as dat

Strona 80 - 76 9650-0801-01 Rev. A

Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 77isolated serial bus through the FPGA. The IBP probes connect to the monitor board throug

Strona 81 - Defibrillator Pacer Module

CHAPTER 6FUNCTIONAL DESCRIPTION78 9650-0801-01 Rev. Amodule contains power supplies, charging and discharge circuitry, pacing circuitry, and the Pads/

Strona 82 - 78 9650-0801-01 Rev. A

Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 79PrinterPropaq MD uses an internal thermal printer of 50 mm paper size that is built into t

Strona 83 - Lithium Ion Battery

CHAPTER 6FUNCTIONAL DESCRIPTION80 9650-0801-01 Rev. A

Strona 84 - 80 9650-0801-01 Rev. A

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 81Appendix AOverviewThis appendix includes the following diagrams:• Propaq MD Electrical Hardware In

Strona 85

82 9650-0801-01 Rev. APropaq MD Electrical Hardware Interconnect Diagram

Strona 86 - 82 9650-0801-01 Rev. A

Propaq MD Signal and Power Interconnect Schematic9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 83Propaq MD Signal and Power Interconnect Schematic

Strona 87

84 9650-0801-01 Rev. A

Strona 88 - 84 9650-0801-01 Rev. A

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 85Appendix BOverviewThis appendix includes simulators and settings that may be used to assess the pe

Strona 89

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 5Chapter 1 Maintenance TestsPropaq MD OverviewThis chapter includes step-by-step procedures for the

Strona 90 - Bio-Tek Instruments BP Pump2

86 9650-0801-01 Rev. ABio-Tek Instruments BP PumpAdult/Pediatric ModeNeonate ModeBio-Tek Instruments BP Pump2Adult/Pediatric ModeNeonate ModeFastBP St

Strona 91 - Clinical Dynamics Smart Arm

Overview9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 87DNI Nevada CuffLinkAdult/Pediatric ModeClinical Dynamics Smart ArmAdult/Pediatric ModeNeon