Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 31
SpO
2
Test
Tools Needed SpO
2
Simulator: Fluke Index 2 or equivalent, Nellcor
TM
SpO
2
finger probe
Test Setup Disconnect any ECG cable or IBP probes from the Propaq MD.
Do This... Observe This... Pass/Fail
1.
Power on the Propaq MD without the
SpO
2
sensor attached.
Verify that the SpO
2
numeric
displays INITIALIZING, changes
to SEARCHING, then goes blank.
2.
Connect the
SpO
2
sensor to the Propaq MD. Verify that SpO
2
numeric
displays SEARCHING.
3.
Set the simulator to 70%
SpO
2
and 60 PPM.
Attach the
SpO
2
sensor to the simulator.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
oo
oo
4.
Set the simulator to 80%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
oo
oo
5.
Set the simulator to 90%
SpO
2
and 60 PPM.
Verify that after a brief period of
initialization, the
SpO
2
numeric
displays an
SpO
2
value between
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
oo
oo